CRYSTA APEX V 700 Series CMM
The 700 models of Mitutoyo's new generation CNC coordinate measuring machines family "CRYSTA-APEX V" offer a high-performance solution designed for small and medium-sized workpieces. Sakarya Bull Metal provides superior accuracy and efficiency in measurement processes with these high-precision systems.
Real-Time Temperature Compensation
One of the most striking features of this series is "real-time temperature compensation." The device compensates for variables such as the measuring table, workpiece, and ambient temperature on the measurement results by sensing these factors.
While traditional CMM systems have to work in constant temperature laboratory environments, this series can work with high accuracy in the range of 16 °C - 26 °C.
In this way, the need for special air conditioning for the production line is reduced; the measurement process becomes more practical and provides savings in time and operator resources.
While traditional CMM systems have to work in constant temperature laboratory environments, this series can work with high accuracy in the range of 16 °C - 26 °C.
In this way, the need for special air conditioning for the production line is reduced; the measurement process becomes more practical and provides savings in time and operator resources.
Confidence in Quality with CMM Inspection Services
The air going into this machine should be as dry as possible. However, if the air is not extremely humid or does not contain high levels of oil vapor, you can directly supply air from the air line to the CMM's maintenance unit. When there is too much moisture or oil vapor in the air, it is recommended to use an air dryer or water drainage unit.
A grounded outlet with a nominal output of 220VAC +10% / -5%, 50/60 Hz should be used to provide power to the system. The power provided must have a sufficient margin exceeding the system's consumption rate.
Voltage: 100V–240VAC +10% / -5%
Frequency: 50/60 Hz
Grounding: Class D or higher grounding is recommended.
Voltage: 100V–240VAC +10% / -5%
Frequency: 50/60 Hz
Grounding: Class D or higher grounding is recommended.
Humidity does not directly affect measurement accuracy, but high relative humidity can cause critical parts of the machine to rust. Electronic components can also be adversely affected. To reduce these risks, ambient humidity should be kept in the range of 55% - 65% as much as possible.
Humidity range in which the device can operate: 40% - 75% RH
Humidity range in which the device can operate: 40% - 75% RH
Maximum amplitude for frequencies below 10 Hz: 2 µm (P-P).
Acceleration for frequencies between 10 Hz and 50 Hz: below 0.004 m/s (0.4 Gal).
When the vibration in the environment exceeds these limits, it is recommended to strengthen the foundation or use optional vibration isolation support.
Acceleration for frequencies between 10 Hz and 50 Hz: below 0.004 m/s (0.4 Gal).
When the vibration in the environment exceeds these limits, it is recommended to strengthen the foundation or use optional vibration isolation support.
The compressed air going into this machine should be as dry and clean as possible. If it is very humid or the oil vapor level is high, it is recommended to use an air dryer and water drainage. This prevents dirt and moisture from the air line from damaging sensitive mechanical and pneumatic components.